IBM DS4700 manual Environmental requirements and specifications, Temperature and humidity

Models: DS4700

1 270
Download 270 pages 52.86 Kb
Page 56
Image 56

Environmental requirements and specifications

This section describes the environmental requirements and specifications for the DS4700 Express Storage Subsystem, including temperature and humidity, altitude, airflow and heat dissipation, shock and vibration requirements, and acoustic noise levels.

Temperature and humidity

Table 13 lists the acceptable temperature and humidity ranges for the DS4700 Express Storage Subsystem when in storage or in transit. Table 14 lists temperature and humidity ranges for the storage subsystem in typical Information Technology (IT) or office environments. Table 15 lists temperature and humidity ranges for the storage subsystem in a Telco/NEBS-3 compliant environment.

Note: Substantial deviations from the suggested operating range, in either direction, if sustained for extended periods of time, will expose the unit to greater risk of failure from external causes.

Important: The recommended operating environment air temperature is 22° C (72° F) or lower.

Table 13. Temperature and humidity requirements for storage subsystem when in storage or in transit

Condition

Parameter

Requirement

 

 

 

Temperature

Storage range

-10° - 65° C (14° - 149° F)

 

 

 

 

Maximum rate of change

15° C (27° F) per hour

 

 

 

 

Transit range

-40° C to 65° C (-40° F to

 

 

149° F)

 

 

 

 

Maximum rate of change

20° C (36° F) per hour

 

 

 

Relative humidity (no

Storage range

8% to 90%

condensation)

 

 

Transit range

5% to 95%

 

 

 

 

 

Maximum dew point

26° C (79° F)

 

 

 

 

Maximum gradient

10% per hour

 

 

 

Table 14. Temperature and humidity requirements for storage subsystem in a typical

Information Technology (IT) or office environment

 

 

 

 

Condition

Parameter

Requirement

 

 

 

Temperature

Operating range

10° - 35° C (50° - 95° F)

 

 

 

 

Maximum rate of change

10° C (18° F) per hour

 

 

 

Relative humidity (no

Operating range

8% to 80%

condensation)

 

 

 

 

 

Table 15. Temperature and humidity requirements for storage subsystem in a Telco/NEBS-3 compliant environment

Condition

Parameter

Requirement

 

 

 

Temperature1

Operating range

5° - 40° C (41° - 104° F)

 

Short-term2

-5° - 50° C (23° - 122° F)

 

 

 

 

Maximum rate of change

30° C (54° F) per hour

 

 

 

24IBM System Storage DS4700 Express Storage Subsystem: Installation, User’s and Maintenance Guide

Page 56
Image 56
IBM DS4700 manual Environmental requirements and specifications, Temperature and humidity, Condition Parameter Requirement