MAINTENANCE | WTC 600 LP / MP |
Replacing the membrane
-Set the pump lift adjustment knob (8/1) to 100 %.
-Unscrew the 4 screws, size 6 mm, at the metering head and remove the metering head.
-Unscrew the membrane to the left.
-Reattach the sealing membrane1), intermediate ring and support disc again.
-Screw in new membrane.
-Bring the push rod to the “end of suction stroke” position by switching the mains voltage on/ off.
-Reassemble metering head and tighten screws over cross (tightening torque 4 Nm).
-Bleed the pump (8/3) and put into operation again.
-After 48 hours of operation, retighten the metering head screws (tightening torque 4 Nm).
4.3Reverse Osmosis Module (RO Module)
4.3.1RO Filter Replacement
The RO filters are to be replaced when
cleaning and disinfection as a measure to1 increase the capacity have been without
success.
Filter replacement:
-Select the switching off operating mode and release the system pressure.
-Loosen the connections (concentrate connection (18/2), drinking water connection (18/1) and supply (18/5)) and unscrew pressure tube (18/4) with tube clamps (19/5) from the frame.
CAUTION Do not loosen the tube clamps because of height adjustment.
-Release securing rings (19/1).
-Unscrew connections (19/2).
-Screw extractor tool into thread of drinking water connection (19/3) and carefully lever out using pipe pliers.
-Remove adapter from lid, clean and insert.
-Take out filter and remove lower lid.
2
3
4
4
5
NOTE Do not interchange lids.
Figure 18
1)Membrane replacement set: Item No. 6.762 172.0