Karcher 600 CD manual 3.2Operating parameters, PRESSURE SWITCH is inactive

Models: 600 CD

1 93
Download 93 pages 15.74 Kb
Page 75
Image 75

WTC 600 LP / MP

APPENDIX

3.2Operating parameters

The control system allows the following operating parameter settings (realization see item 4). The resulting behaviours are described in the chapter operating conditions. The values in brackets are the standard values after system reset (see item 4.4.4).

LIM

function:

Permeate limit value used to switch the unit off in

 

 

case limit is exceeded or if a DISPOSAL process

 

 

is to be finished. If this value is exceeded during

 

 

PRODUCTION, RINSE or CONCENTRATE

 

 

DISPLACEMENT/FLUSH for the time defined for

 

 

t-delay LIM, an ALARM MESSAGE will be

 

 

activated.

 

right of access:

user/technician

 

possible setting: meas. range 0.5-50µS/cm (40 µS/cm)

 

 

Attention: the set value must be above aaLIM

aaLIM

function:

Permeate limit value activating an ALARM

 

 

MESSAGE in the display and on/off of the

 

 

malfunction LED if this value is exceeded for the

 

 

time defined for t-delay aaLIM

 

right of access:

user/technician

 

possible setting:

0.5-50µS/cm (10 µS/cm)

 

 

Attention: the set value must be lower than LIM

t-delay LIM

function:

Time delay for wrong permeate limit value

 

 

(see above)

 

right of access:

technician

 

possible setting:

0-99 min (1 min)

t-delay aaLIM

function:

Time delay for advance alarm in case of wrong

 

 

permeate limit value (see above)

 

right of access:

technician

 

possible setting:

0-99 min (5 min)

t-delay lack of pressure

function:

Time delay to release PRESSURE FAILURE if

 

 

PRESSURE SWITCH is inactive.

 

right of access:

technician

 

possible setting:

0-99s (1s)

t-delay at start

function:

Period of time of permanent pressure at start of

 

 

 

operation after opening of the inlet valve before

 

 

 

the pumps begin to work.

 

 

right of access:

technician

 

 

possible setting:

0-9.9s (0.5s)

t-pressure available

function:

Period of time until PRESURE FAILURE is

 

 

 

released if pumps do not start after opening of

 

 

 

inlet valve.

 

 

right of access:

technician

 

 

possible setting:

0-99s (5s)

t rinse

function:

Duration of a DISCONTINUOUS RINSE.

 

 

right of access:

technician

 

 

possible setting:

0-99 min (5s)

 

 

 

 

Service Handbook 08.2004

 

75

Page 75
Image 75
Karcher 600 CD manual 3.2Operating parameters, PRESSURE SWITCH is inactive