WTC 600 LP / MP | APPENDIX |
4.4.2Calibration menu
The calibration menu is used to adjust the sensors. The following functions are available:
Menu point: | Function: | ||||
1 Cy permeate | Calibration of permeate sensor. If the menu point '1 Cy permeate' is | ||||
| chosen, a submenu appears with two items: '1 min. Cy' and2'max. Cy'. | ||||
| '1 min Cy': please remove the Cy sensor and press the . key. | ||||
| '2 max Cy' | ||||
| Important!!! The temperature must be calibrated before calibration of the | ||||
| conductivity. | ||||
| Connect the Cy sensor. Now the current permeate conductivity has to be | ||||
| determined by means of a standard meter, and this value has to be | ||||
| entered by numerical editing. The entered value is equated with the | ||||
| current conductivity after actuation of the Í |
| key. | ||
|
| ||||
2 temp.permeate | Calibration of permeate temperature sensor. | ||||
| For that purpose the current permeate temperature has to be defined | ||||
| with a standard meter, and this value has to be entered by means of | ||||
| numerical editing. The entered value is equated with the current | ||||
| temperature when the Í |
| key is then actuated. | ||
|
| ||||
|
| ||||
3 calibr. reset | This menu point allows to reset adjustment of the two measuring inputs | ||||
| to the internal standard values. | ||||
| ATTENTION !! Internal calibration is not a correct adjustment. |
Service Handbook 08.2004 | 87 |