APPENDIX | WTC 600 LP / MP |
4.4.1Main menu
The details of the different menu points of the main menu are described below.
Menu point: | Function: |
1 Ackn. failure | This menu point allows to acknowledge an operating failure |
| (the only possibility if ps acknowledgement = with password). |
2 Operat. mode | This menu point allows to change the operating mode by means of a |
| selection. For safety purposes, selection only can be made between |
| STOP and OPERATION as well as STOP and DISINFECTION |
| respectively (direct |
| is not possible). |
2 Cy permeate | Permanent display of current permeate conductivity. To leave this menu |
| point press one of the keys. |
4 Temp. permeate | Permanent display of current permeat temperature. To leave this menu |
| point press one of the keys. |
5 Hour meter | |
6 Cy limits | Numerical entry of LIM and aaLIM. |
7 Optional alarm | Setting (selection) of ps LIM high |
8 Calibration | This menu point assures branching to the calibration menu (see item |
| 4.4.3). The new calibration will be stored if the entered data are |
| acknowledged before leaving the calibration menu. |
9 Setting | This menu point assures branching to the setting menu (see item 4.4.4). |
| The modification will be stored of the entered data are acknowledged |
| before leaving the setting menu. |
10 Diagnosis | This menu point assures branching to the diagnosis menu (see item |
| 4.4.5). At the beginning of the diagnosis, all actuators of the unit are |
| switched off. The sytem returns to the previous status if diagnosis is |
| completed. |
86 | Service Handbook 08.2004 |