Maintaining the Equipment

Cleaning the Light Bar

If artifacts appear on your scanned images, clean the light bar.

IMPORTANT: Wear white cotton gloves for this procedure to prevent oils and dirt from being absorbed into the microfiber cleaning cloth.

1.Fold a dry, untreated microfiber cleaning cloth so the edge of the cloth is about the thickness of the light bar.

NOTE: You can use an e-wipe, manufactured by Photographic Solutions, Inc., instead of a microfiber cleaning cloth.

2.Gently move the edge of the microfiber cloth or e-wipedown the length of the light bar one time.

3.Calibrate the scanner.

4.If an artifact remains after cleaning the light bar, check the light bar with a magnifying glass for damage, such as a scratch or a chip. If there is damage to the light bar, contact your Kodak representative for service.

Replacement Procedures

Replacing the Lamp

White cotton gloves are required for this procedure.

One spare lamp is provided with the scanner.

IMPORTANT: Use only the GE ELC Lamp with the HR Film Scanner (see

Appendix B: Ordering Accessories and Supplies).

It is not necessary to power down the scanner before changing the lamp.

lamp-and-filter- access door

lift door open from here

P192_0016HC

1. Open the lamp-and-filter-access door.