leads

Maintaining the Equipment

lamp

Hot Surface symbol

socket

thumbscrews

lamp cover

P192_0023GC

2.Loosen the thumbscrews on the lamp cover and pull the cover straight out.

When you remove the cover, the safety switch is released, removing power to the lamp socket.

CAUTION: Wait at least five minutes for the lamp to cool.

3.Push down the lamp release lever.

4.Remove the old lamp from the lamp socket.

5.Fully seat the new lamp in the socket; polarity is not important.

CAUTION: In the next step, to ensure correct alignment with the light bar, it is extremely important to push the lamp to the back of the lamp holder with the leads facing to the left. Improper alignment affects the operation of the scanner.

6.With the socket leads facing left, reinstall the lamp in the lamp holder.

7.Reinstall the lamp cover and tighten the thumbscrews.

Power is reapplied to the lamp socket and the lamp should light.

8.Calibrate the scanner (see Calibrating the Scanner on page 3-5).

May 2002

4-3

Page 45
Image 45
Kodak 6B7198 manual Leads