Maintaining the Equipment

6.Replace the balance filters:

a.Remove the 3 screws.

b.Remove the filter retainer.

c.Remove the old filters and place the new filters in the filter holder.

The filters are coated on both sides. It is not important which side faces up or down.

d.Secure the filters with the retainer and the 3 screws.

7.Install the filter assembly in the scanner with the balance-filter selection handle in the middle.

8.Reinstall the lamp cover, tighten the thumbscrews, and select a filter position.

9.Calibrate the scanner (see Calibrating the Scanner on page 3-5).

Replacing the (Square) IR Filter

IMPORTANT: The IR filter is made of glass and is expensive to replace. It is not included in the product warranty or service contract. Use care when handling the IR filter.

Wear white cotton gloves for this procedure.

It is not necessary to power down the scanner before changing the IR filter.

1.Open the lamp-and-filter-access door (see page 4-4for the location of the lamp-and-filter-access door).

IR filter (under balance- filter assembly)

red handle

2.Loosen the thumbscrews on the lamp cover and pull the cover straight out.

When you remove the cover, the safety switch is released, removing power to the lamp socket.