6. Assembly

6.3 Tube and Eyepieces

Note:

For fluorescence applications, install the fluorescence illuminator* first p. 21.

If available, the analyzer* (10.1) must be inserted into the stand. This requires that the guide key engages in the guide pin (10.2).

To mount the analyzer, the analyzer mount TL* 20 mm or 60 mm can also be placed between stand and tube.

An intermediate tube pole* with a switchable analyzer (on/off) and Bertrand lens is also available as an option.

The tube is mounted to the stand either directly or with the use of intermediate modules*.

Loosen the clamping screw (11.1) on the stand

Insert the tube in the circular receptacle (dovetail ring)

Retighten the clamping screw (11.1)

The eyepieces are inserted into the eyepiece tubes on the tube

6.4 Objectives

Always only use Leica objectives of tube length ¥ (infinity)! The standard thread is M25. The objectives should be arranged so that the magnification increases when the objective nosepiece is rotated counter-clockwise.

!

Attention:

Lower the specimen stage as far as possible before assembling the objectives. Close vacant threads in the nosepiece with dust protection caps!

Fig. 10 Analyzer mounting

1Analyzer

2

Orientation pin and guiding notch

Fig. 11 Fastening the tube

3

Clamping screw

1 Clamping screw

1

19

Page 19
Image 19