6. Assembly
a
b
3
Hg 50
Xe 75
1
4
2
3
1
2
5
• Insert the lamp mount, with the burner in-
stalled, into the lamp housing and tighten it
with the screws (15.8)
• Close the lamp housing and retighten the
screws
• Place the lamp housing in the incident light
lamp housing receptacle (17.1) and fasten it
with the clamping screw on the side
• Connect the lamp housing to the external
power supply
Fig. 17 Mounting the 106z lamp housing
1Lamp housing receptacle
1
Fig. 16 a-c Lamp mounts for gas discharge lamps
1Upper clamping system
2Lower clamping system
3Cooling element
4Nipple of the mercury 50 burner,
5Dust cover of the mercury 75 burner
c
Hg 100
3
1