11. Trouble shooting

Problem

Cause/Remedy

 

 

Fluorescence: The lamp does not illuminate immediately upon being switched on.

￿The external power supply must be switched on repeatedly

￿Hot Hg lamps should cool down before switching on again

Focus

The specimen cannot be brought into focus.

Darkfield

￿Use the correct immersion medium

￿Lay the specimen with the cover glass to- wards the top

￿Make sure that the cover glass thickness is correct and that it suits the indication on the objective

No definite DF contrast is possible.

￿Be sure that a DF objective is being used

 

￿The objective aperture setting is too high

 

(maximum 0.75/1.10); if necessary, reduce the

 

objective aperture using the iris diaphragm on

 

the objective

 

￿Check the condenser centering

 

￿Open the aperture diaphragm completely

The image is unevenly or not uniformly illuminated.

￿The magnification is too weak. Use a higher magnification

Undesirable stray light.

￿Clean the specimen and neighboring lenses

 

(p. 54)

Polarization

 

 

 

No polarization contrast is possible.

￿Bring the polarizer and analyzer into cross po- sition until they reach maximum darkness (without specimen) (p. 45)