6. Assembly

6.6Components for Fluorescence Applications* 6.6.1 Fluorescence illuminator*

The fluorescence illuminator is mounted before the tube. It is fastened in place with the side clamping screw (13.1).

6.6.2 106z Lamp Housing*

Caution!

Light sources pose a potential irradiation risk (glare, UV-radiation, IR-radiation). Therefore, lamps have to be operated in closed housings.

During assembly, always unplug the power supply unit of the 106z lamp housing from its socket.

During assembly work on xenon burners, al- ways wear the supplied protective gloves and face protection (fig. 14) (risk of explosion).

Never touch the glass parts of the burner with bare hands.

Never look directly into the beam path (blinding hazard).

The lamp housing 106z is used with various gas discharge lamps.

Caution!

Make sure to follow the instructions and safety notes of the lamp supplier.

Before changing lamps allow at least 30 mins for cooling down!

Inserting the Gas Discharge Lamps* (Hg and Xe) into the 106z Lamp Housing

Hg and Xe lamps are powered by separate supply units.

Read the separate instruction manual provided with these supply units.

Fig. 13 Assembly of fluorescence illuminator

1Clamping screw

1

Fig. 14

Protective gloves and mask

21

Page 21
Image 21
Leica DM1000 LED manual 2 106z Lamp Housing, Hg and Xe lamps are powered by separate supply units