8. Operation

8.4 Tubes

Note:

Close any unused tube openings, as otherwise stray light can interfere with observation.

Adjusting the Viewing Distance

Adjust the viewing distance of the eyepieces so that a congruent total image is seen (fig. 40).

Adjusting the Viewing Angle

For the HC LVB 0/4/4 and HC -/0/4 ergonomy tubes*, the viewing angle can be adjusted by tilting the binocular viewer

Ergotube (long, swivelable): 0° - 35°

Ergotube (short, swivelable): 7.5° - 32.5°

For the AET22 and EDT22 ergotubes*, the viewing angle can be adjusted by tilting the binocular viewer in the range of 5° - 32° (fig. 41)

Adjusting the Eyepiece Section to the Arm Length

On the AET22 tube, the eyepieces can be extended up to 30 mm (fig. 41)

Beam Splitting in Photo Tubes*

EDT22 tube:

The beam splitting between the observation and documentation outputs has a definite presetting (50:50).

BDT P 25 tube:

The beam splitting is set manually by pulling out a control bar.

Control Bar

 

 

 

 

 

Observation

Photo

VIS

 

 

 

 

 

100%

0%

50/50

 

 

 

 

 

 

 

 

50%

50%

PHOTO

 

 

 

 

 

0%

100%

 

 

 

 

 

 

 

Fig. 41 With AET22* tube individual adjustments

Fig. 40 Tube setting

Personal viewing distance of the eyepieces

37

Page 37
Image 37
Leica DM1000 LED manual Operation Tubes, Adjusting the Viewing Distance, Adjusting the Viewing Angle