4.2 Parameters

Parameter

Parameter Description

 

Data

number

 

Type

 

 

 

1

Load Cell mV/V (1, 2 or 3. 0 = unknown).

Float

2

Units: 0=g, 1=kg, 2=ton, 3=oz, 4=lb.

Float

3

Full Load Cell(s) Capacity

Float

4

Maximum Applied Capacity

Float

5

Resolution Index (0-17)

 

 

Resolution Index is actually the index to an array of 18

 

 

defined values(0-17) like that:

 

 

0.0001,

0.0002, 0.0005,

 

 

0.001,

0.002,

0.005,

 

 

0.01,

0.02,

0.05,

Float

 

0.1,

0.2,

0.5,

 

 

1,

2,

5,

 

 

10,

20,

50.

 

 

E.g., if Resolution Index=8, then system resolution =

 

 

 

 

0.05.

 

8

Calibration Date formatted as "MMDDYY".

Float

 

So, 10107 <= Calibration Date <= 123199.

 

 

11

Calibration Time formatted as "HHMM".

Float

 

So, 0 <= Calibration Time <=2359

 

 

20

Auto transmit interval (3-52734, integer).

 

 

How many internal adc updates (52734 Hz) there are

 

 

between auto transmissions.

 

 

 

 

This gives a theoretical reading rate from 17578 per

 

 

sec to 1 per sec. (i.e. rate=52734/P20). Practically, the

 

 

actual rate for low values of P20 is usually less than

Float

 

the theoretical rate. (“P20” stands for “Parameter

 

 

 

 

 

#20”.) Note that this rate relates to USB

 

 

 

communication; upon using a serial port, the rate is far

 

 

smaller.

 

 

 

 

Becomes effective only after a system reset (either

 

 

power off/on or using the ‘S’ command).

 

34

Page 35
Image 35