Variables Table

Category

Variable Name

 

Description

Get

Set

 

Calibration_Name

Name of calibration

V

 

 

Calibration_Date

Calibration date (MMDDYY).

V

 

 

Calibration_Time

Calibration time

V

 

 

(HHMM, e.g., 1545).

 

 

 

 

 

 

Unit

 

Weighing unit:

V

 

 

ton, kg, g, lb or oz.

 

 

 

 

 

 

 

Weighing resolution:

 

 

 

 

0.0001, 0.0002, 0.0005, 0.001,

 

 

Calibration

 

0.002, 0.005,

 

 

 

Resolution

0.01, 0.02, 0.05,

 

V

 

Info

 

 

 

0.1, 0.2, 0.5,

 

 

 

 

 

 

 

 

 

 

1, 2, 5,

 

 

 

 

 

 

 

10, 20 or 50.

 

 

 

Full_Capacity

Full capacity of the load cell.

V

 

 

Maximum_Load

Maximum applied load on the

V

 

 

 

load cell.

 

 

 

 

 

 

 

Load_Cell_Output

Output of the load cell:

V

 

 

 

0 = Unknown

 

 

 

1, 2 or 3 = 1, 2 or 3 mV/V.

 

 

 

 

There are two filters. The first filter is

 

 

 

 

 

 

 

 

 

 

 

 

Filter1

basically a 1st level moving average

V

*

 

filter of size Filter1 (2-256). Then

 

 

depending on Decimator (1-1000),

 

 

 

 

every Decimator-th result from the 1st

 

 

 

 

 

 

 

 

level filter is put through another

 

 

 

Filter2

moving average filter of size Filter2

V

*

Filtering

(2-256), which is the second filter.

 

 

 

 

 

 

 

 

 

* Setting these three variables causes

 

 

 

 

 

 

 

 

a board reset, so it is a slow function.

 

 

 

 

Therefore, in order to set all of them

 

 

 

Decimator

there is one more option: the method

V

*

 

Set_Filtering , which works faster

 

 

 

 

 

 

than three individual activations of

 

 

 

 

the Set_Variable method.

 

 

39