Chapter 1 Calibration
34
Calibration
WARNI NG
Class 2 laser light. Do not stare into the laser beam or a
reflected image of the laser beam.
WARNI NG
Using controls, making adjustments, or performing
procedures other than those specified herein may result in
hazardous radiation exposure.
The validator should be calibrated when it is first installed or when
the printer displays the Calibration warningmessage. A special
calibration bar code card is included with the validator package.
1. Makesure the printer is on and the scanning beam is on. If the
beam is off, press the RESET button for less than 2 seconds to
turn on the beam.
2. Raisethe media cover. Open the pivoting deck by rotating the
deck lock lever fully clockwise. The pivoting deck will swing
upward, exposing the bottom of the printhead assembly.
3. Ifthe printer has the cutter option, put the cutter in the open
(down) position.
4. Removethe calibration bar code card from its prote ctive sleeve
and slide it into the open printhead and under the scanning
beam. (Figure 17.)
5. Makesure the bar code is under the scanning beam and the
CALIBRATION light is on.
6. Pressand hold the RESET button on the validator until the
CALIBRATION light begins flashing.
Table2. RESET Button Settings
Time RESET Button Is Depressed Result
Less than 2 seconds Turns on scanning beam
More than 4 but less than 6 seconds Validator goes into calibration mode
More than 6 seconds Resetsvalidator to factory calibration and
turns off scanning beam