Defining Validator Options

6.Press to select the desired value.

7.Press and at the same time to lock the key, then press PAUSE to take the printer offline.

8.Press PAUSE again to put the printer online.

Setting The Symbol Contrast

Symbol Contrast represents the contrast between the bars and spaces in the bar code. The bar code must have a certain level of contrast to be recognized by the validator. When you enable the Symbol Contrast parameter, the bar code contrast is included as part of the pass/fail criteria.

.

1. Press .. to take the printer offline and place the printer in Menu mode.

2. If necessary, press and at the same time to unlock the key.

.

3. If necessary, press .. until “VALIDATOR” displays.

4. Press until “Symbol Contrast” displays.

5. Press + or to scroll through the choices: Enable or Disable. The default is Enable.

6. Press to select the desired value.

7. Press and at the same time to lock the key, then press PAUSE to take the printer offline.

8. Press PAUSE again to put the printer online.

Checking Quiet Zones

Quiet zones are the white spaces surrounding the bar code. Each bar code requires a minimum quiet zone distance in order for the bar code to be scanned properly.

NOTE: The validator requires a minimum distance of 1/2” or 20 times the minimum element width (x-dimension), whichever is greater, between bar codes.

When the Quiet Zones parameter is enabled, then the bar code quiet zone is included as part of the pass/fail criteria.

.

1. Press .. to take the printer offline and place the printer in Menu mode.