| INSTALLATION 3 |
| SITRANS FX300 | |
|
|
|
3.6.2 Installing devices in flange design
Figure 3-12: Installing devices in flange design
1Gasket
2Bolts with fixing nuts
• Use bolts and fastening nuts 2 to attach the measuring device to one side of the flange.
•While doing so, insert the gaskets 1 between measuring sensor and flange and align them.
•Check that the gasket is concentric and that it is not protruding into the pipe
•Install the gasket, bolts and fastening nuts on the other side of the flange.
•Align the measuring device and the gaskets so they are concentric.
•Now tighten all nuts bit by bit alternately across the diagonal.
06/2009 - SFIDK.PS.050.F2.02 A5E02100423 | www.siemens.com/flow | 29 |