OPERATION 6

 

SITRANS FX300

 

 

 

Totalizer / pulse output

The base units for the totalizer and the pulse output are m3 for volume,

m3 norm. for standard volume and kg for mass.

The unit and the conversion factor can be changed in menu item 3.2.8 "Tot.Conf.". User defined units (User Def.) can be selected and entered. The conversion factor of the unit must always be entered based on the base unit.

The maximum pulse output frequency is 0.5 Hz. To ensure that the pulse output does not exceed

0.5Hz, the conversion factor of the totalizer must be selected and set accordingly. The pulse output represents an exact copy of the whole number value of the internal totalizer.

Max. pulses per hour

fmax. 0.5 Hz

1 pulse-break 2 seconds

Max. pulses / hour = 1800 pulses

Example

Flow Qmax.

Factor A1 Coef.

Pulses / litre

Pulses / h

Comment

 

 

 

 

 

5.6 m3/h

0.001

1 pulse / 1 litre

5600

not possible

 

0.01

1 pulse / 10 litres

560

560 < 1800 = ok.

 

 

 

 

 

Menu items

1.1.3

Maes.Inst

Volume / Standard volume / Mass

 

 

 

1.1.4

Max. Flow

Unit (User Def.) / Text / A1 Coeff. / Max. Flow / Flow Displ

 

 

 

3.2.5

Function P

Yes / No

 

 

 

3.2.7

Totalizer

Tot. on / Tot. off

 

 

 

3.2.8

Tot. Conf.

User def. / Unit Text / A1 Coeff. / Preset Value / Reset / Display

 

 

 

6.2.6 Measures in the event of faulty indications

If the indications on the display or the responses to keypad commands are faulty, you have to do a hardware reset. Switch the power supply OFF and ON again.

06/2009 - SFIDK.PS.050.F2.02 A5E02100423

www.siemens.com/flow

43