6 OPERATION |
|
|
SITRANS FX300 |
| |
|
|
|
6.5.8 Menu item Setup (firmware version - gas)
Level | Designation | Selection / entry | Explanation | ||
|
|
|
|
|
|
3.1 |
|
|
|
| Set display |
|
|
|
|
|
|
3.1.1 | Error message |
|
|
| Display error |
| → |
|
|
|
|
| Yes | ↑... |
| Shows error messages in measuring mode in plain text | |
|
|
| |||
|
|
|
|
| alternating with the measured values |
|
|
|
|
| |
|
| No | ↑...^ | Flashing cursor in the top left corner of the display | |
|
|
|
|
| indicates errors. |
3.1.2 | Cycle.Disp. |
|
|
| Display measured values |
| → |
|
|
|
|
| Yes | ↑... |
| Cyclically display measured values in measuring mode | |
|
|
| |||
|
|
|
|
| (6 s intervals) |
|
|
|
|
| |
|
| No | ↑...^ | Do not display measured values cyclically | |
|
|
|
|
|
|
3.1.3 | Code 1 |
|
|
| Set security query 1 (menu) |
| → |
|
|
|
|
| Yes | ↑... |
| Activate access code: | |
|
|
| |||
|
|
|
|
| →→→↑↑↑^^^ |
|
|
|
|
| |
|
| No | ↑...^ | Deactivate access code | |
|
|
|
|
|
|
|
|
|
|
|
|
3.2 |
|
|
|
| I/O settings |
|
|
|
|
|
|
3.2.1 |
|
|
| Current output | |
| → |
|
|
|
|
| Current output | ||||
|
| ||||
|
|
|
|
| |
|
| ↑... | Current output | ||
|
|
|
| ||
|
| Current output | |||
|
|
|
|
|
|
3.2.5 | Function P |
|
|
| Pulse output |
| → |
|
|
|
|
| Yes | ↑... |
| Activate pulse output | |
|
|
| |||
|
|
|
|
| |
|
| No | ↑...^ | Deactivate pulse output | |
|
|
|
|
|
|
3.2.7 | Totalizer |
|
|
| Totalizer |
| → |
|
|
|
|
| Tot. on | ↑... | Start totalizer | ||
|
| ||||
|
|
|
|
| |
|
| Tot. off | ↑...^ | Stop totalizer (value is held) | |
|
|
|
|
|
|
3.2.8 | Tot. Conf. |
|
|
| Configure totalizer |
| → |
|
|
|
|
|
|
|
|
|
|
| (available if Maes.Inst = | m3 |
|
| Select totalizer unit for volume flow measurement |
| Volume, see menu item | Unit ↑...^ |
| ||
| 1.1.3) |
|
|
|
|
| 0000000000 | Enter preset totalizer value | |||
|
| ||||
|
| m3 ↑…↑...↑...^ |
| ||
|
| Reset Yes / | Reset totalizer | ||
|
| Reset No ↑...^ | Do not reset totalizer/ | ||
|
|
|
| ||
|
| Disp. off / | Do not display totalizer | ||
|
| Disp. on | ↑...^ | Display totalizer / | |
|
|
|
|
| |
| (available if Maes.Inst = | m3 norm |
| Select totalizer unit for standard volume flow | |
| Norm. Vol., see menu | Unit | ↑...^ | measurement | |
| item 1.1.3 |
|
|
|
|
| 0000000000 | Enter preset totalizer value | |||
|
| ||||
|
| m3 norm ↑…↑...↑...^ |
| ||
|
| Reset yes / | Reset totalizer | ||
|
| Reset no | ↑...^ | Do not reset totalizer/ | |
|
|
|
| ||
|
| Disp. off / | Do not display totalizer | ||
|
| Disp. on | ↑...^ | Display totalizer / | |
|
|
|
|
|
|