6 OPERATION

 

 

 

 

 

 

 

 

 

 

 

SITRANS FX300

 

 

 

 

 

 

 

 

 

 

 

 

Level

Designation

Selection / entry

Explanation

 

 

 

 

 

 

 

 

 

3.3.3

HART TV

 

 

 

HART tertiary variable

 

 

 

 

 

 

 

 

 

 

Temp. ...

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Pressure ...

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Density

...^

 

 

 

 

 

 

 

 

 

 

3.3.4

HART 4V

 

 

 

HART quaternary variable

 

 

 

 

 

 

 

 

 

 

Temp. ...

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Pressure ...

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Density

...^

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

3.4

 

 

 

 

 

Set fluid and medium

 

 

 

 

 

 

 

 

 

3.4.1

Fluid 1

 

 

 

Fluid type

 

 

 

 

 

 

 

 

 

 

Steam ^

 

Steam

 

 

 

 

 

 

 

 

 

 

 

 

 

 

3.4.2

Medium

 

 

 

Process medium

 

 

 

 

 

 

 

 

 

 

Sat.Steam ...

Saturated steam

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Sup.Steam ...^

Overheated steam

 

 

 

 

 

 

 

 

 

3.4.5

Dry. Fact.

 

 

 

Enter portion of dryness in saturated steam

 

 

 

 

 

 

 

 

 

 

0000001.00

Factor = 1 - %water mass

 

 

 

(available if Medium =

 

 

0.85 TO 1

...

 

 

 

 

Sat.Steam, see menu item

 

 

 

 

3.4.2)

 

...^

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1 Density check

 

 

 

 

 

 

 

 

 

 

 

 

 

 

3.5

 

 

 

 

 

Pressure, temperature and density

 

 

 

 

 

 

 

 

 

3.5.1

T-sensor

 

 

 

Internal temperature sensor

 

 

 

 

 

 

 

 

 

 

No ...

 

No temperature sensor available

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Yes

...^

Temperature sensor available

 

 

 

 

 

 

 

 

 

3.5.2

P-sensor

 

 

 

Pressure sensor

 

 

 

 

 

 

 

 

 

 

Intern ...

 

Internal pressure sensor

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

---

...^

No pressure sensor available

 

 

 

 

 

 

 

 

 

3.5.3

Sat.P/T

 

 

 

Density calculation for saturated steam

 

 

 

 

 

 

 

 

 

 

Sat. Temp. ...

Density calculation for saturated steam from saturated

 

 

 

(available if Medium =

 

 

 

 

 

temperature

 

 

 

Sat.Steam, see menu item

 

 

 

 

 

3.4.2)

 

Sat. Pres. ...^

Density calculation for saturated steam from saturated

 

 

 

 

 

 

 

 

pressure

 

 

 

 

 

 

 

 

3.5.4

Temp. Opr.

 

 

 

Operating temperature

 

 

 

 

 

 

 

 

 

 

°C

 

 

Set temperature unit

 

 

 

(for saturated steam with

...^

 

 

 

Sat. Temp. or Sup.Steam,

Unit

 

 

 

 

 

see menu items 3.4.2 and

0000000.0

 

Operating temperature

 

 

 

3.5.3)

 

 

 

 

 

°C ^

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Disp. on /

 

Display temperature /

 

 

 

 

 

Disp. off ...^

Do not display temperature

 

3.5.5

Pres. Opr.

 

 

 

Operating pressure

 

 

 

(available for Sat.Steam

 

 

 

 

 

 

 

Pa

 

 

Set pressure unit

 

 

 

with Sat. P or Sup.Steam,

 

 

 

 

Unit ...^

 

 

 

 

 

see menu items 3.4.2 and

 

 

 

 

 

3.5.3)

 

0000000.0

 

Operating pressure

 

 

 

 

 

Pa ^

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Disp. on /

 

Display pressure /

 

 

 

 

 

Disp. off ...^

Do not display pressure

 

 

 

 

 

 

 

 

 

 

56

www.siemens.com/flow

06/2009 - SFIDK.PS.050.F2.02 A5E02100423

Page 56
Image 56
Siemens FX300 manual Hart tertiary variable Temp. ↑ Pressure ↑ Density