6OPERATION
6.4 Error messages
SITRANS FX300
Error message | Cause | Measure | |
|
|
| |
No Signal | No signal from vortex amplifier | Check connector | |
|
| In the event of measuring sensor | |
|
| problems contact service. | |
|
|
| |
Low freq | Sample frequency too low | Contact service. | |
|
|
| |
High Freq | Sample frequency too high | Contact service. | |
|
|
| |
Low Flow | Flow rate less than the set minimum | Signal converter continues to display the | |
| flow rate qmin | current flow rate but the accuracy of the | |
|
| measurement may be affected. | |
Q too high | Flow rate higher than the set maximum | Corrective measures depend on the | |
| flow rate qmax | application! | |
|
| If the flow rate exceeds the maximum | |
|
| value, the measuring sensor may be | |
|
| physically damaged. | |
|
|
| |
Inv. Config. | Invalid configuration data in FRAM | Check entire configuration! | |
| (permanent memory) | Contact service if error message | |
|
| persists. | |
|
|
| |
Amp. Fail | Error in the | Contact service. | |
|
|
| |
Chk. Instal. | Quality of the vortex signal too poor | 1. Check qmin | |
|
| 2. Excessive pipe vibration and faulty | |
|
| flow profile | |
|
|
| |
Low Signal | Amplitude of vortex signal is too low | 1. qmin ok? | |
|
| 2. If qmin OK, contact service | |
Hi. Signal | Amplitude of vortex signal is too high | Occurs with media of high density | |
|
| 1. Check qmax | |
|
| 2.If qmax OK, contact service | |
L.Temp.Phy | Operating temperature lower than | Take corrective measures within | |
| specified | process. | |
|
|
| |
H.Temp.Phy | Operating temperature higher than | Take corrective measures within | |
| specified | process as quickly as possible, | |
|
| otherwise both the measuring sensor | |
Hi.P.Phys | Pressure higher than specified | ||
and the signal converter may be | |||
|
| damaged. | |
|
|
| |
Tsens Shrt. | Temperature sensor short circuit | Shows error at temperature sensor! | |
|
| Contact service. | |
Tsens Open | Open temperature sensor | ||
| |||
|
|
| |
P. Sen. Fail | Faulty pressure sensor | Displays error at pressure sensor! | |
|
| Contact service. | |
Table |
|