SCPI Commands for the VX4101A
3–24 VX4101A MultiPaq Instrument User Manual
SYSTem:RQUeue:SNUMber:STATe?
Returns the SNUMber state of 0|1|ON|OFF when the VX4101A is in Asynchro-
nous Mode.
SYSTem:SNUMber?
This query returns the serial number of the VX4101A.
SYSTem:TIMeout?
This query returns the maximum amount of time a query is allowed to take when
the VX4101A is in Synchronous Mode.
SYSTem:VERSion?
Returns the firmware version of the instrument in <major>.<minor>.<sub-mi-
nor> form.
SYSTem:ERRor?
Command Response/description
SYST:ERR? –113,“Undefined header; Command not found; XXX\0D”
SYST:ERR? 0,“No error”
SYSTem:LANGuage?
Command Response
*RST
SYST:LANG? SYNC
SYST:LANG ASYN
SYST:LANG? ASYN
SYSTem:OPTions?
Command Response
SYST:OPT? NOOPT
SYSTem:OPTions:DESCription
Command Response
SYST:OPT? 1C
SYST:OPT:DESC? “1C” Counter 500 MHz option