Using the Digital Output
2–38 VX4101A MultiPaq Instrument User Manual
Outputing One 32-Bit Word
This procedure will show you how to continuously output a 32-bit word at a
specified frequency with the Digital Output.
1. Select the Digital Output:
INST:SEL DIGO
2. Load the first memory location:
TRACe 1,<data>
3. Load the sample rate at the specified frequency:
TRACe:SRATe <sample output frequency>
4. Set the trace length:
TRACe:Points 1
5. Specify continuous output:
INIT:CONTinuous ON
6. Begin the sampling operation:
TRIGger:IMMEDiate