SCPI Commands for the Counter
3–60 VX4101A MultiPaq Instrument User Manual
correction factors for the Counter attenuation and gain settings. This command
requires an input calibration source set to the <input voltage> values below
(±0.1%).
Table 3–3: Input Calibration Source Settings
<input voltage>
+50.0
+20
+10
+5.0
+2.0
+1.0
+0.5
+0.2
+0.1
+0.05
–50.0
–20
–10
–5.0
–2.0
–1.0
–0.5
–0.2
–0.1
–0.05
[<channel>] specifies the channel being calibrated, and can be specified as either
1 or 2. If [<channel>] is not specified, channel one is assumed.
CALibrate[<channel>]:ZERO
The CALibrate[<channel>]:ZERO command performs offset corrections for each
of the gain and attenuation settings of the Counter. This command requires an
input calibration source set to 0.0 V ±0.001 V.
[<channel>] specifies the channel being calibrated, and can be specified as either
1 or 2. If [<channel>] is not specified, channel one is assumed.