Status and Event Reporting System
4–12 VX4101A MultiPaq Instrument User Manual
STATus : OPERation : PTRansition


Instrument
32,767
0–32767
N/A
Sets the Operational Positive Transition Filter for the currently selected
instrument. Setting a bit in this filter latches zero to one transitions of the
corresponding bit in the Operational Condition Register into in the Operational
Event Register.
Queries the Operational Positive Transition Filter for the currently selected
instrument. Response is in NR1 format with a range of 0-32767.
Command Syntax
Query Syntax
Command Class
*RST Value
Limits
Related Commands
Command Description
Query Response