SCPI Commands for the SurePath Modules
VX4101A MultiPaq Instrument User Manual 3–263
INSTrument:ABORt
Places active instrument in the IDLE state and ceases any measurement or other
instrument activity in progress. The instrument configuration is unchanged. A
subsequent INIT command will cause the instrument to re-start the same type of
measurement.
If the instrument is in Asynchronous mode, this command can be sent while
a query is in progress and the measurement will be aborted.
If the instrument is in Synchronous mode, this command will be queued
while a query is in progress.
NOTE. After abort, no more measurements are taken.
INSTrument:RESet
Resets the currently selected instrument without affecting other instruments. The
instrument returns to its *RST state. The instrument remains selected.
INSTrument:ABORt
Command Response
INST:SEL DMM
CONF:RES
CONF? “:SCAL:RES 3e+08,180000”
INST:RES
INST:SEL? DMM
CONF? “:SCAL:VOLT:DC 300,0.001”
INSTrument:RESet
See the TRIGger([: SEQuence1] |:STARt) [:LAYer]:IMMediate command for an
example.
Command Description
Examples