SCPI Commands for the SurePath Modules
VX4101A MultiPaq Instrument User Manual 3–289
Valid command for VX4330, VX4350 and VX4380 SurePathmodules
N/A
Parameter Value
<count> 1
<delay in seconds> 0 seconds (pass-through)
<triggers to count> 0 triggers (pass-through)
<source> IMMEDIATE
Parameter Value
<count> 1–65535
<delay in seconds> 0=pass through, 1 ms–65.536 ms in 1 ms steps
<triggers to count> 0 = pass through
1 – 65,536 triggers
<source> BUS,TTLGRG0, TTLTRG1, TTLTRG2, TTLTRG3, TTL TRG4,
TTLTRG5, TTLTRG6, TTLTRG7, COMMAND0, COMMAND1,
COMMAND2, COMMAND3, COMMAND4, SUREPATH, DMM,
COUNTER, CTR_EXTARM, DAC, DIGI, DIGO
ABORt,
[ROUTe:]SCAN
TRIGger([:SEQuence]|:STARt)[:LAYer]:COUNt <count>
Specifies the number of times the instrument will enter the waiting for trigger
state prior to returning to the idle state. After an initiate, the instrument will enter
the wait for trigger state until a trigger is received, at which time the trigger
count will be decremented. The selected operation will take place and if the
trigger count is not zero, the instrument will re-enter the wait for trigger state.
TRIGger([:SEQuence]|:STARt)[:LAYer]:DELay <delay in seconds>
Specifies a time delay to occur after receipt of a trigger command before actually
triggering. If the selected trigger source is fixed, this command will have no
effect on the instrument triggering. This command always zeros the event count
delay, so specifying a delay of zero places the trigger subsystem in pass-through
mode. In this mode, the instrument triggers immediately upon receipt of a trigger
command.
SurePath Module
Query Syntax
*RST Value
Limits
Related Commands
Description