SCPI Commands for the Digital Input
3–136 VX4101A MultiPaq Instrument User Manual
Command/Query Response/Description
SENS:PSEL:DIB DIS-
ABLE,#hf0 Disables bits 5 – 8 as input bits.
SENSe:PSELect? #h0000000f
Queries all bit states.
SENSe:MODE <FORMatted|UNFormatted>
Command/Query Response/Description
SENS:MODE FORM Sets the data returned to be formatted.
SENS:MODE? FORMATTED
Queries the mode for the data format.
SENSe:SRATe?
Command/Query Response/Description
SENSE:SRATE 48000.0 Sets the sample rate to 48 kHz.
SENS:SRAT? 48000.0
Queries the sample rate.
SENSE:SRATE 24100.0 Sets the sample rate to 24.1 kHz
SENS:SRAT? 24100.0
Queries the sample rate.
SENSe:THReshold?
Command/Query Response/Description
SENSE:THRESHOLD 10 Sets the threshold.
SENS:THR? 10.000
Queries the threshold.