Appendix G: Calibration
VX4101A MultiPaq Instrument User Manual G–5
This part of the procedure calibrates the DMM for 0.3 VDC with 10 GW
impedance.
33. Program the VX4101A as follows:
conf:dc .3;init
34. Apply 0.0 VDC ±1 mVDC
35. Program the VX4101A as follows:
cal:val 0
36. Apply 0.290 VDC ±5.3 mVDC
37. Program the VX4101A as follows:
cal:val .29 (or any other desired value)
This part of the procedure calibrates the DMM for 3 VDC with 10 GW imped-
ance.
38. Program the VX4101A as follows:
conf:dc 3;init
39. Apply 0.0 VDC ±4 mVDC
40. Program the VX4101A as follows:
cal:val 0
41. Apply 2.90 VDC ±35 mVDC
42. Program the VX4101A as follows:
cal:val 2.9 (or any other desired value)
DC Range 0.3 VDC with
10 GW Impedance
DC Range 3 VDC with
10 GW Impedance