Using the Counter
VX4101A MultiPaq Instrument User Manual 2–31
This will return the time interval from the first edge on channel one to the
first edge on channel two that is at least 1 msec later.
12. Set the aperture to minimum:
SENS:APER MIN
13. Initiate the configured measurement:
INIT
14. Place the result of the measurement in the output buffer so you can read it:
FETch?
15. Set the time interval with delay by events:
SENSe12:FUNCtion "TINT:DEL:EVEN"
16. Set the number of events to 10:
SENSe:TINTerval:DELay:EVENTS 10
This will return the time interval from the first edge of channel one to the
tenth edge on channel two.
17. Initiate the configured measurement:
INIT
18. Place the results of the measurement in the output buffer so you can read
them:
FETC?