Appendix G: Calibration
VX4101A MultiPaq Instrument User Manual G–11
38. Program the VX4101A as follows:
cal:val 0
39. Apply 10 MW ±2000 W
40. Program the VX4101A as follows:
cal:val 10e6 (or any other desired value)
This part of the procedure calibrates the DMM for 300 MW resistance.
41. Program the VX4101A as follows:
conf:fres 300e6;init
42. Apply 0.0 W ±1600 W
43. Program the VX4101A as follows:
cal:val 0
44. Apply 100 MW ±200 kW
45. Program the VX4101A as follows:
cal:val 100e6 (or any other desired value)
NOTE. 2-wire resistance W ranges use the above calibration information. A
separate 2-wire W calibration (other than at 30 W) is not required.
Resistance 300 MW