Appendix

Models 5000B

 

 

 

 

 

 

 

 

 

Typical Analytical Performance Specifications:

(will vary per application)

Accuracy:

±2% of full scale or better, application dependent

Noise:

Less than ±1%

Diurnal:

Less than 1% per 20oF (10oC (source/detector

 

dependent)

Sample Cell:

Stainless steel with Sapphire window

 

standard. Other materials available.

Cell Length:

.01 to 10 inches

Flow Rate:

50 to 1500 cc/min

Light Source:

TungstenMini-Lamp, optional Quartz Halogen

Sensitivity:

.015 to 3 absorbance units.

Reproducibility:

+/- 1% of scale or better

Filter Wavelength:

750nm to 5 microns, application dependant

Sample Pressure:

Sapphire window 250 psi

Response time:

Programmable

Other Specifications:

Ambient temp range: 32-122°F (0-50°C)

Internal calibration: Optical span flag/zero and span calibration

NOTE: 5000B/5020B (sample temperature sensing and output correction)

Power Rating: 115/230 VAC, 50/60 Hz, Approx 200 watts

Readout Device: Digital, 2 line alphanumeric vacuum fluorescent display, (VFD)

Analog Output Signal: 0-1 V (standard), 4-20mA isolated

Area Classification

 

Non-hazardous:Model 5000B

 

Hazardous / Explosion-proof:

Model 5020

5000B, 5020B Purge options:X, Y, Z purge, Cenelec IIB

AND H2 (PENDING) FOR ZONE I OR CLASS I, DIV I, B, C, D

AREAS