1 Introduction

Model 5000B

 

 

 

 

 

 

 

 

 

Normally closed

Normally open

Moving contact

Normally open

Moving contact

Figure

2-6: Types of Relay Contacts The connectors are:

Threshold Alarm 1: • Can be configured as high (actuates when concentration is above threshold), or low (actuates when concentration is below thresh old).

Can be configured as fail-safe or non-fail-safe.

Can be configured as latching or nonlatching.

Can be configured out (defeated).

Threshold Alarm 2:

• Can be configured as high (actuates when concen-

 

tration is above threshold), or low (actuates when

 

concentration is below threshold).

 

• Can be configured as fail-safe or non-fail-safe.

 

• Can be configured as latching or nonlatching.

 

• Can be configured out (defeated).

System Alarm:

Actuates when DC power supplied to circuits is

 

unacceptable in one or more parameters. Permanently

 

configured as fail-safe and latching. Cannot be de-

 

feated. Actuates if self test fails.

 

To reset a System Alarm during installation, discon-

 

nect power to the instrument and then reconnect it

 

Further detail can be found in chapter 3, section 3-5.

Digital Remote Cal Inputs

Remote Zero and Span Inputs: The REMOTE SPAN and RE- MOTE ZERO inputs are on the DIGITAL INPUT terminal block. They

2-6: Part I