3 Operation/Control Unit

Model 5000B

 

 

 

 

 

 

 

 

 

3.7The Analyze Function

Normally,allofthefunctionsautomaticallyswitchbacktotheAnalyze function when they have completed their assigned operations. Pressing the Escape button in many cases also switches the analyzer back to the Analyze function. Alternatively, you can press the Analyze button at any time to return to analyzing your sample.

The Analyze function screen shows the impurity concentration and the application fluides in the first line, and the range in the second line. In the lower right corner, the abbreviation Anlz indicates that the analyzer is in the Analyze mode. If there is an * before the Anlz, it indicates that the range is linearized.

1.95ppm SO2 R1:Ø —10 *Anlz

If the concentration detected is overrange, the first line of the display blinks continuously.

3.8Programming

CAUTION: The programming functions of the Set Range and Curve Algorithm screens are configured at the factory to the users application specification. These functions should only be reprogrammed by trained, qualified personnel.

To program, you must:

1.Enter the password, if you are using the analyzer’s password protection capability.

2.Connect a computer or computer terminal capable of sending an RS-232 signal to the analyzer RS-232 connector. (See chapter 2 Installation for details). Send the rp command to the analyzer.

3.Press the System button to start the System function.

DIG_FILT SELF-TEST

PWD LOGOUT MORE

Use the < > arrow keys to blink MORE, then press Enter.

3-28 Part I

Teledyne Analytical Instruments

Page 56
Image 56
Teledyne M5000B operating instructions Analyze Function