Photometric Analyzer | Operation /Control Unit 3 | |
|
|
|
|
|
|
|
|
|
Note: Before setting up an
Note: If you require highly accurate
To setup an
Choose System from the Function buttons. TheVFD will display five subfunctions.
DIG_FILT
PWD LOGOUT MORE
Select MORE and press the Enter Key
AUTOCAL | FILSOL | HOLD |
MORE |
|
Use < > arrows to blink
ZERO in Ød Øh off
SPAN in Ød Øh off
Press < > arrows to blink ZERO (or SPAN), then press Enter again. (You won’t be able to set OFF to ON if a zero interval is entered.) A Span Every ... (or Zero Every ...) screen appears.
Zero schedule: OFF
Day: Ød Hour: Øh
Use DÑ arrows to set an interval value, then use < > arrows to move to the
To turn ON the SPAN and/or ZERO cycles (to activate
If instrument is turned off, the next time the instrument is powered, the instrument will automatically perform a calibration cycle after 3 minutes of entering the sample mode if AUTOCAL functions were on prior to shut down.