Teledyne M5000B operating instructions System Self-Diagnostic Test

Models: M5000B

1 106
Download 106 pages 40.63 Kb
Page 37
Image 37

Photometric Analyzer

Operation /Control Unit 3

 

 

 

 

 

 

 

 

 

3.3.4System Self-Diagnostic Test

The Model 5000BF has a built-in self-diagnostic testing routine. Pre- programmed signals are sent through the power supply, output board, preamp board and sensor circuit. The return signal is analyzed, and at the end of the test the status of each function is displayed on the screen, either as OK or as a number between 1 and 1024. (See System Self Diagnostic Test in chapter 4 for number code.) If any of the functions fails, the System Alarm is tripped.

Note: The sensor will always show failed unless Zero fluid is present in the sampling cell at the time of the SELF-TEST.

The self diagnostics are run automatically by the analyzer whenever the instrument is turned on, but the test can also be run by the operator at will. To initiate a self diagnostic test during operation:

Press the System button to start the System function.

DIG_FILT SELF-TEST

PWD LOGOUT MORE

Use the < > arrow keys again to move the blinking to the SELF–TEST and press Enter. The screen will follow the running of the diagnostic.

RUNNING DIAGNOSTIC

Testing Preamp — Cell

When the testing is complete, the results are displayed.

Power: OK Analog: OK

Cell: 2 Preamp: 3

The module is functioning properly if it is followed by OK. A number indicates a problem in a specific area of the instrument. Refer to Chapter 5 Maintenance and Troubleshooting for number-code information. The results screen alternates for a time with:

Press Any Key

To Continue...

Then the analyzer returns to the initial System screen.

Teledyne Analytical Instruments

Part I 3-9

Page 37
Image 37
Teledyne M5000B operating instructions System Self-Diagnostic Test