3 Operation/Control Unit

Model 5000B

 

 

 

 

 

 

 

 

 

• Latching?Ltch:Y/N(Yes/No).

To define the setpoint, use the < > arrow keys to move the blinking over to AL1: ####. Then use the DÑ arrow keys to change the number. Holding down the key speeds up the incrementing or decrementing.

To set the other parameters use the < > arrow keys to move the blinking over to the desired parameter. Then use the DÑ arrow keys to change the parameter.

Once the parameters for alarm 1 have been set, press Alarms again, and repeat this procedure for alarm 2 (AL2).

To reset a latched alarm, go to Dft– and then press either D two times or Ñ two times. (Toggle it to Y and then back to N.)

–OR

Go to Ltch– and then press either D two times or Ñ two times. (Toggle it to N and back to Y.)

3.6The Range Select Function

The Range function allows you to manually select the concentration range of analysis (MANUAL), or to select automatic range switching (AUTO).

In the MANUAL screen, you are further allowed to define the high and low (concentration) limits of each Range, and select a single, fixed range to run.

CAUTION: If this is a linearized application, the new range must be within the limits previously programmed using the System function, if linearization is to apply throughout the range. Furthermore, if the limits are too small a part (approx 10 % or less) of the originally linearized range, the linearization will be compromised.

3.6.1Manual (Select/Define Range) Screen

The Manual range-switching mode allows you to select a single, fixed analysis range. It then allows you to redefine the upper and lower limits, for the range.

Press Range key to start the Range function.

3-24 Part I

Teledyne Analytical Instruments

Page 52
Image 52
Teledyne M5000B operating instructions Range Select Function, Manual Select/Define Range Screen