3 Operation/Control Unit

Model 5000B

 

 

 

 

 

 

 

 

 

Use the DÑ keys to toggle between AUTO and MAN span settling. Stop when MAN appears, blinking, on the display. Press Enter to move to the next screen.

Span Val: 2Ø.ØØ % <ENT> To begin span

Use the < > arrow keys to toggle between the span concentration value and the units field (%/ppm). Use the DÑ arrow keys change the value and/ or the units, as necessary. When you have set the concentration of the span fluid you are using, press Enter to begin the Span calibration.

Press Enter to enter the span value into the system and begin the span calibration.

Once the span has begun, the microprocessor samples the output at a predetermined rate. It calculates the difference between successive samplings and displays this difference as Slope on the screen. It takes several seconds for the first Slope value to display. Slope indicates rate of change of the Span reading. It is a sensitive indicator of stability.

####.##% SO2 Slope=#.### Span

When the Span value displayed on the screen is sufficiently stable, press Enter. (Generally, when the Span reading changes by 1 % or less of the range being calibrated for a period of ten minutes it is sufficiently stable.) Once Enter is pressed, the Span reading changes to the correct value. The instrument then automatically enters the Analyze function.

3.4.3Offset Function

This software when installed in the 5000BF instruments provides a way to enter an offset on the zero operation of the analyser. This is useful when the instrument is zeroed in some inert fluid such as Nitrogen or Argon, but the background fluid of the process is different. If the background fluid of the process is different than the zero calibration fluid being used, the reading will have an offset that will be constant throughout its working range. Thus, the need to provide an offset when the instrument is being zeroed.

How to access the offset function:

To access this function, the instrument zero mode must be entered by pushing the Zero key on the front panel of the control unit. The VFD display will show the following menu selection:

3-20 Part I

Teledyne Analytical Instruments

Page 48
Image 48
Teledyne M5000B operating instructions Offset Function, How to access the offset function