Teledyne M5000B operating instructions TeledyneAnalyticalInstruments4-13

Models: M5000B

1 106
Download 106 pages 40.63 Kb
Page 81
Image 81

Operations/Analysis Unit 4

always be corrected using the calibration zero and/or span features.

SUMMARY

The zero and/or span calibration of 5000 series photometric Teledyne analyzers can be quickly verified manually by using Zero and/or calibration zero offsets and/or span flags.

Refer to typical Teledyne Analytical 1475 sample system below, 3-way valves 1,2,+ 3 below are used to check internal calibration using zero and/or span filters with N2.

CALIBRATION OF H2O IN SOLVENTS

EXAMPLE

Based upon our experience when suppling photometric NIR analyzers for 0-100ppm water in solvents like EDC, TDI, benzene and the like, the following can be incorporated for the customers pertinent solvent.

1Customer is responsible for introducing a reproducible dry zero fluid, typically (-75deg C) dewpoint Nitrogen or reproducible dry EDC or TDI etc., within +/- 1 ppm H2O (for accuracy of +/- 1 % full scale).

2When customer supplies a reproducible dry TDI zero fluid: A span flag option is available to verify instrument full scale sensitivity. This optical flag may be (manually or automatically) introduced into the light path within the temperature controlled (within +/- 1 deg C) sample cell compartment. This optical span flag ( when in this temperature stable environment simulates a very stable upscale ppm H2O reading on top of the DRY reproducible TDI zero background.

3Conversely, a zero flag and/or offset is (manually or automatically) switched into the light path of the analyzer after drying out the sample cell of process fluid. Consult your manual for the factor calibration values obtained.

Actual concentration

Zero check

5 ppm water/TDI or

0 ppm water in N2

Span Check mole-sieve or dried sample (could be from process) 5ppm water in TDI 85ppm water in TDI (grab sample from process)

5000/5020 series analyzer meter

5ppm water in the 5020 meter or output. Check.

5ppm water on the 5020 meter or output. with zero flag offset introduction.

delta 0ppm water or stability check on zero verified.

5ppm water on the 5020 meter or output. Zero verified. 85 ppm water as determined by Karl Fischer titration. (lab technique must be verified for precision. (sampling device, Karl Fischer titrator and operator technique)

TeledyneAnalyticalInstruments4–13

A Business Unit of Teledyne Electronic Technologies

Page 81
Image 81
Teledyne M5000B operating instructions TeledyneAnalyticalInstruments4-13