Operations/Analysis Unit 4
3.If the analyzer section has been equipped with the automatic zero SOL option, the zero solenoid device in the accessory sampling system should have energized the instant power was established. The device (or devices) should be energized because the mode switch has been preset to the ZERO position.
4.If the test procedure was normal, the devices should have been seen or heard to operate as described by personnel located at the analysis section installation site, and no further check need be made at this time. If operation is not as described, refer to Troubleshooting in Chapter 5.0.
5.Check analyzer operation by completing the calibration procedure described below in Calibration after allowing the instrument to warm up. The electrical circuits take from 30 to 40 minutes to stabilize.
NOTE: It will take at least four hours for the analyzer to completely stabilize with respect to temperature.