Teledyne M5000B operating instructions Part II Maintenance/Analysis Unit

Models: M5000B

1 106
Download 106 pages 40.63 Kb
Page 95
Image 95

Part II: Maintenance/Analysis Unit

Maintenance 5

 

 

 

 

 

 

 

 

 

Theprocedureispurelymechanicalandconsistsofadjustingtheamountof light passing through either the measuringor reference filter, never both. Screens (wire mesh) of varying density are used for this operation and are part of the small tookkitaccompanyingtheinstrument.

1.Observe the oscilloscope and judge if optical balancing is needed. When the difference is less than 1 volt, balancing is not required. The tallest of the two peaks should be adjusted to 8 or 9 volts with the gain control R2 on the detector PC board. When this cannot be done because both peaks are too short or too long, search for screens mounted in the light path, usually located in a holder on the light pipe which interconnects the detector and sample module, and remove or add screens, as necessary.

2.When balancing is needed, identify the peaks as outlined under Section

3.For example, if the reference peak is the shorter one, stop the filter wheel with your hand and see if screens are located behind the reference filter. The reference filter is identified by the letter “R” engraved on the filter wheel.

4.If screens are found, remove them after taking the filter wheel off the shaft with the special Allen wrench supplied in the tool kit.

5.After removal of the screens and remounting the filter, mount the filter wheel back on the shaft. Position it correctly on the shaft by lining up the two paint marks on shaft and wheel.

6.Turn on the instrument and observe the balance on the oscillo- scope.

a.If the reference peak is now too tall, remove the filter wheel and add a screen of lesser density behind the reference filter. Repeat this procedure until the peaks are within 1 volt of each other.

b.If the measuring peak is equal to or within 1 volt of thereference peak, the system is optically balanced and ready for calibration.

c.If the peak is still too short, repeat the procedure, but thistime put a screen behind the measuring filter to shorten its peak.

Teledyne Analytical Instruments

Part II: 5-5

Page 95
Image 95
Teledyne M5000B operating instructions Part II Maintenance/Analysis Unit