Part I: Control Unit/Analysis Unit

Maintenance 3

 

 

 

 

 

 

 

 

 

Preamp

0 OK

>0 means that some of the ten gains of the amplifier have a large offset. The number is a code that tells which of the ten gtains have the large offset.

Cell (detector)

0 OK

>0 means that the analyzer failed to balance the measuring and the reference signals, the larger the number the farther off they were.4.3 Major Internal Components

The major components in the Control Unit are shown in Figure 4-3.

Figure 4-3: Control Unit Major Internal Components

Teledyne Analytical Instruments

Part I: 4-3