Photometric Analyzer

Part I: Control Unit

 

 

 

 

 

 

 

 

 

accept 0 V (OFF) or 24 V dc (ON) for remote control of calibration (See Remote Calibration Protocol below.)

Zero:

Floating input. 5 to 24 V input across the + and – terminals

 

puts the analyzer into the ZERO mode. Either side may be

 

grounded at the source of the signal. 0 to 1 volt across the

 

terminals allows ZERO mode to terminate when done. A

 

synchronous signal must open and close the external zero

 

valve appropriately. See Remote Probe Connector at end of

 

section 3.3. (With the -C option, the internal valves automati-

 

cally operate synchronously).

Span:

Floating input. 5 to 24 V input across the + and – terminals

 

puts the analyzer into the SPAN mode. Either side may be

 

grounded at the source of the signal. 0 to 1 volt across the

 

terminals allows SPAN mode to terminate when done. A

 

synchronous signal must open and close the external span

 

valve appropriately. See Remote Probe Connector at end of

 

section 3.3. (With the -C option, the internal valves automati-

 

cally operate synchronously.)

Cal Contact: This relay contact is closed while analyzer is spanning and/or zeroing. (See Remote Calibration Protocol below.)

Remote Calibration Protocol: To properly time the Digital Remote Cal Inputs to the Model 5000B Analyzer, the customer's controller must monitor the Cal Relay Contact.

When the contact is OPEN, the analyzer is analyzing, the Remote Cal Inputs are being polled, and a zero or span command can be sent.

When the contact is CLOSED, the analyzer is already calibrating. It will ignore your request to calibrate, and it will not remember that request.

Once a zero or span command is sent, and acknowledged (contact closes), release it. If the command is continued until after the zero or span is complete, the calibration will repeat and the Cal Relay Contact (CRC) will close again.

For example:

1)Test the CRC. When the CRC is open, Send a zero command until the CRC closes (The CRC will quickly close.)

2)When the CRC closes, remove the zero command.

3)When CRC opens again, send a span command until the CRC closes. (The CRC will quickly close.)

4)When the CRC closes, remove the span command.

Teledyne Analytical Instruments

Part I: 2-7

Page 25
Image 25
Teledyne M5000B operating instructions Part I