SENSe:]FUNCtion:TOTalize

[SENSe:]FUNCtion:TOTalize <ch_list> sets the SENSe function to TOTalize for channels in <ch_list>. See “Totalize Positive or Negative Edge State Changes” on page 15.

Parameters

 

 

 

 

 

 

 

Parameter

Parameter

Range of

Default

 

 

Name

Type

Values

Units

 

 

 

 

 

 

 

 

ch_list

string

100 - 163

none

 

 

 

 

 

 

 

 

 

 

 

 

 

Comments The totalize function counts rising edges of digital transitions at Frequency/Totalize SCP channels. The counter is 24 bits wide and can count up to 16,777,215.

The SENS:TOT:RESET:MODE command controls which events will reset the counter.

If the channels specified are not on a Frequency/Totalize SCP, an error will be generated.

Related Commands: SENS:TOT:RESET:MODE,

INPUT:POLARITY

*RST Condition: SENS:FUNC:COND and INP:POL NORM for all digital SCP channels.

Send with VXIplug&play Function: hpe14XX_cmd(...)

Usage See “Totalize Positive or Negative Edge State Changes” on page 15.

SENS:FUNC:TOT (@134)

channel 34 is a totalizer

Agilent E1538A Enhanced Frequency/Totalize/PWM SCP 55

Page 59
Image 59
VXI VT1538A user manual SENSeFUNCtionTOTalize