VXI VT1538A user manual SOURcePULSePERiod?, SOURcePULSeWIDTh, Parameters

Models: VT1538A

1 84
Download 84 pages 24.07 Kb
Page 75
Image 75
SOURce:PULSe:PERiod?

Usage SOUR:PULS:PER .005,(@140)

set PWM pulse train to 200 Hz

 

on channel 40

SOURce:PULSe:PERiod?

SOURce:PULSe:PERiod? (@<channel>) returns the fixed pulse period value on the pulse width modulated pulse channel in <channel>.

Parameters

 

 

 

 

 

 

 

Parameter

Parameter

Range of

Default

 

 

Name

Type

Values

Units

 

 

 

 

 

 

 

 

channel

string

100 - 163

none

 

 

 

 

 

 

 

 

 

 

 

 

 

Comments If the channels specified are not on a Frequency/Totalize SCP, an error will be generated.

Returned Value: numeric period. The type is float32.

Send with VXIplug&play Function: hpe14XX_cmdReal64_Q(...)

SOURce:PULSe:WIDTh

SOURce:PULSe:WIDTh <pulse_width>,(@<ch_list>) sets the fixed pulse width value on the frequency modulated pulse channels in <ch_list>.

Parameters

 

 

 

 

 

 

 

Parameter

Parameter

Range of

Default

 

 

Name

Type

Values

Units

 

 

 

 

 

 

 

 

pulse_width

numeric (float32)

7.87E-6 to 7.8125E-3

seconds

 

 

 

 

(238.4E-9 resolution)

 

 

 

 

 

 

 

 

 

ch_list

string

100 - 163

none

 

 

 

 

 

 

 

 

 

 

 

 

 

Comments If the channels specified are not on a Frequency/Totalize SCP, an error will be generated.

*RST Condition: SOUR:FM:STATE OFF and SOUR:PULM:STATE OFF

Related Commands: SOUR:PULM:STATE, SOUR:PULS:POLarity

The variable frequency control for this channel is provided by the algorithm language. When the algorithm executes an assignment statement to this channel, the value assigned will be the frequency setting. For example:

O143 = 2000 /* set channel 43 to 2KHz */

Agilent E1538A Enhanced Frequency/Totalize/PWM SCP 71

Page 75
Image 75
VXI VT1538A user manual SOURcePULSePERiod?, SOURcePULSeWIDTh, Parameters, Send with VXIplug&play Function hpe14XXcmdReal64Q