Totalizer

Capacity

24 bits or 16,777,215

 

Minimum Pulse Width

500nS

 

Frequency Range

0-100 KHz

Frequency Measurement

Gate Time ( taperture )

1 mSec to 1 Second, resolution

1

 

 

f--i--n-

 

 

 

 

 

Range

1

to 100, 000

 

 

 

t--a---pe-----r--t--ur-----e

 

 

 

 

 

 

Accuracy

.01%

 

 

Resolution (Hz)

fi n pu t

 

 

 

----------------------------------------------------

 

 

 

ta pe rt ure

× 4.194MHz

 

 

Minimum Pulse Width

500 nS

 

Period Measurement

Gate Time ( taperture )

40 μSec to 4 Second

 

Range (SENS:PER:RANGE=1)

10μSec to ta pe rtu re

(1 sec max)

 

(SENS:PER:RANGE=4)

40μSec to tap ert u re

(4 sec max)

 

Accuracy

.01%

 

 

Resolution

.2384μSec

 

Minimum Pulse Width

500 nSec

Time-out Mechanism

Programmable 4mSec to 60Sec

(SENS:FREQ:LIM:LOW & SENS:PER:LIM:UPP)

Aperture Time

The aperture time is the time allowed to average multiple period and frequency measurements

Function

Aperture Range

 

 

Frequency

1mSec to 1 Sec

Period (2 ranges)

1mSec to 1 Sec (SENS:PER:RANGE 1)

 

4mSec to 4 Sec (SENS:PER:RANGE 4)

 

 

78 Agilent E1538A Enhanced Frequency/Totalize/PWM SCP

Page 82
Image 82
VXI VT1538A user manual Frequency Measurement