Totalizer | Capacity | 24 bits or 16,777,215 |
| Minimum Pulse Width | 500nS |
| Frequency Range |
Frequency Measurement | Gate Time ( taperture ) | 1 mSec to 1 Second, resolution | 1 | |
|
| |||
|
|
|
| |
| Range | 1 | to 100, 000 |
|
|
|
| ||
|
|
|
| |
| Accuracy | .01% |
| |
| Resolution (Hz) | fi n pu t |
| |
|
|
| ||
|
| ta pe rt ure | × 4.194MHz |
|
| Minimum Pulse Width | 500 nS |
|
Period Measurement | Gate Time ( taperture ) | 40 μSec to 4 Second | |
| Range (SENS:PER:RANGE=1) | 10μSec to ta pe rtu re | (1 sec max) |
| (SENS:PER:RANGE=4) | 40μSec to tap ert u re | (4 sec max) |
| Accuracy | .01% |
|
| Resolution | .2384μSec | |
| Minimum Pulse Width | 500 nSec |
Programmable 4mSec to 60Sec |
(SENS:FREQ:LIM:LOW & SENS:PER:LIM:UPP)
Aperture Time
The aperture time is the time allowed to average multiple period and frequency measurements
Function | Aperture Range |
|
|
Frequency | 1mSec to 1 Sec |
Period (2 ranges) | 1mSec to 1 Sec (SENS:PER:RANGE 1) |
| 4mSec to 4 Sec (SENS:PER:RANGE 4) |
|
|