*RST Condition: is 1

If SENSe:PERiod:RANGe = 1.0

then settings conflict if 1.0sec < APERture < 4.0sec. If SENSe:PERiod:RANGe = 4.0

then settings conflict if 0.01sec <= APERture < 0.04sec

The E1415/19/22 driver will report these Settings Conflicts at INITiate time with the following error message:

3129, Incompatible Aperture and Range values,SCP[x]

Related Commands: SENSe:PERiod:NPERiods

*RST Condition: is 1

.Send with VXIplug&play Function: hpe14XX_cmd(...)

[SENSe:]PERiod:RANGe[:UPPer]?

[SENSe:]PERiod:RANGe[:UPPer]? <channel> returns the current setting of upper period limit.

Parameters

 

 

 

 

 

 

 

Parameter

Parameter

Range of

Default

 

 

Name

Type

Values

Units

 

 

 

 

 

 

 

 

channel

string

100 - 163

none

 

 

 

 

 

 

 

 

 

 

 

 

 

Comments <channel> must be single channel on an E1538A SCP.

RETURNS: float32, configured upper period time range for given ch. Response is in seconds. Returned value either 1 or 4

.Send with VXIplug&play Function: hpe14XX_cmdReal64_Q(...)

[SENSe:]TOTalize:RESet:MODE

[SENSe:]TOTalize:RESet:MODE <select>,<ch_list> sets the mode for

resetting totalizer channels in <ch_list>.

Parameters

 

 

 

 

 

 

 

Parameter

Parameter

Range of

Default

 

 

Name

Type

Values

Units

 

 

 

 

 

 

 

 

select

discrete (string)

INIT TRIGger

seconds

 

 

 

 

 

 

 

 

ch_list

string

100 - 163

none

 

 

 

 

 

 

 

 

 

 

 

 

 

Comments In the INIT mode the total is reset only when the INITiate command is executed. In the TRIGger mode the total is reset every time a new scan