VXI VT1538A user manual SENSeTOTalizeRESetMODE?, SOURceFMSTATe, Parameters, Comments

Models: VT1538A

1 84
Download 84 pages 24.07 Kb
Page 67
Image 67
[SENSe:]TOTalize:RESet:MODE?
Channel must specify a single channel.

is triggered.

If the channels specified are not on a Frequency/Totalize SCP, an error will be generated.

Related Commands: SENS:FUNC:TOT, INPUT:POLARITY

*RST Condition: SENS:TOT:RESET:MODE INIT

Send with VXIplug&play Function: hpe14XX_cmd(...)

Usage SENS:TOT:RESET:MODE TRIG,(@134)

totalizer at channel 34 resets at

 

each trigger event

[SENSe:]TOTalize:RESet:MODE?

[SENSe:]TOTalize:RESet:MODE? <channel> returns the reset mode for

the totalizer channel in <channel>.

Parameters

 

Parameter

Parameter

Range of

Default

 

 

Name

Type

Values

Units

 

 

 

 

 

 

 

 

channel

string

100 - 163

none

 

 

 

 

 

 

 

 

 

 

 

 

 

Comments

If the channel specified is not on a frequency/totalize SCP, an error will be generated.

Returned Value: returns INIT or TRIG. The type is string.

Send with VXIplug&play Function: hpe14XX_cmdString_Q(...)

SOURce:FM[:STATe]

SOURce:FM[:STATe] <enable>,(@<ch_list>) enables the Frequency

Modulated mode for a PULSe channel.

Parameters

 

 

 

 

 

 

 

Parameter

Parameter

Range of

Default

 

 

Name

Type

Values

Units

 

 

 

 

 

 

 

 

enable

boolean (uint16)

1 0 ON OFF

none

 

 

 

 

 

 

 

 

ch_list

string

100 - 163

none

 

 

 

 

 

 

 

 

 

 

 

 

 

Comments This command is coupled with the SOURce:PULM:STATE command. If the FM state is ON then the PULM state is OFF. If the PULM state

Agilent E1538A Enhanced Frequency/Totalize/PWM SCP 63

Page 67
Image 67
VXI VT1538A SENSeTOTalizeRESetMODE?, SOURceFMSTATe, Parameters, Comments, Send with VXIplug&play Function hpe14XXcmd