VXI VT1538A user manual SENSePERiodNPERiods?, SENSePERiodRANGeUPPer, Parameters, Comments

Models: VT1538A

1 84
Download 84 pages 24.07 Kb
Page 65
Image 65
SENSe:PERiod:NPERiods?

SENSe:PERiod:NPERiods?

SENSe:PERiod:NPERiods? (@<channel>) returns the number of

signal periods the SCP will measure and average to calculate the signal period.

Parameters

 

 

 

 

 

 

 

Parameter

Parameter

Range of

Default

 

 

Name

Type

Values

Units

 

 

 

 

 

 

 

 

channel

string

100 - 163

none

 

 

 

 

 

 

 

 

 

 

 

 

 

Comments

 

 

 

 

<channel> must be single channel on an E1538A SCP.

RETURNS: uint32 value which is the current period count configured.

Related Commands: SENSe:PERiod:NPERiods

*RST condition: SENSe:PERiod:NPERiods = 1

.Send with VXIplug&play Function: hpe14XX_cmdInt32_Q(...)

[SENSe:]PERiod:RANGe[:UPPer]

[SENSe:]PERiod:RANGe[:UPPer] <range>,<ch_list> can extend the range of period measurement from the default 1 second maximum to a 4 second maximum.

Parameters

 

Parameter

Parameter

Range of

Default

 

 

Name

Type

Values

Units

 

 

 

 

 

 

 

 

range

numeric (float32)

1 4 MAXimum MINimum

Sec

 

 

 

 

 

 

 

 

ch_list

string

100 - 163

none

 

 

 

 

 

 

 

 

 

 

 

 

 

Comments <ch_list> must be channels on an E1538A SCP.

Range 1.0 sec provides period measurement in the range: 10usec - 1sec

Range 4.0 sec provides period measurement in the range: 40usec - 4sec

MINimum = 1.0, MAXimum = 4.0

Note 1538As have a possible settings conflict:

Agilent E1538A Enhanced Frequency/Totalize/PWM SCP 61

Page 65
Image 65
VXI VT1538A SENSePERiodNPERiods?, SENSePERiodRANGeUPPer, Send with VXIplug&play Function hpe14XXcmdInt32Q, Parameters